New Requirements
(DOSECTRL-1) The DoseControl software product installer. The Installer puts the DoseControl system software, modules, and hardware drivers on the PC where the installer is launched.
(DOSECTRL-14) The DoseControl software product installer contains an upgrade workflow. If the user has an older version of DoseControl on the PC, the Installer will provide a method for the user to upgrade to the new version of the software.
(DOSECTRL-21) The software installer and associated modules are signed with a secure Code Signing process.
(DOSECTRL-188) The DoseControl software product installer contains an previous version removal workflow. If the user has an older version of DoseControl on the PC, the Installer will provide a method for the user to remove the previous version from the PC.
Resolved Anomalies (Bug fixes)
(DOSECTRL-120) Users upgrading from DoseControl v1.x to v2.x, the Report Summary screen could display a false alert for ‘Dose outside calibration range’.
(DOSECTRL-204) For users with multiple DoseControl computers connected to a single application database, there is a possibility of a database deadlock during concurrent updates, with no potential for lost data but causing a software crash without an error message.
(DOSECTRL-2222) In DoseControl v2.0, when the dosimetry system calibration is configured with Dose Units (Gy), the dose units did not function as expected on the report measurement screen.
Known Anomalies (non-critical):
(DOSECTRL-80) If the user has ‘Edit Thickness’ permission enabled in the Dosimeter Type configuration, the user with permission is allowed to change the dosimeter thickness on the measurement screen.
(DOSECTRL-81) While the modification of the Calibration configuration End Date event is recorded accurately in the audit trail, the event is identified as GxP event and should have the user enter a comment to be recorded in the audit trail.
(DOSECTRL-82) Audit trail should record a change to the report's storage location.