A summary of the highlights in this release.
New Features
-
[GEX-2174] - Calibration Output Validation Report function was added to the calibration configuration screen. Users can now verify the output of calibrations over a range of fixed absorbance values by generating this PDF and comparing with a dose table or other method to validate the actual curve output.
-
[GEX-2196] - Filter Visible Reports on Search Screen by Pathway
-
[GEX-2181 & GEX-2183] - Product Specification Module now functions for non-integrated systems - Catalog Number drop down can be used in headers to select a product and import the related product specifications from the module into a dosimetry report.
Improvements
This release contains the following improvements:
-
[GEX-2165] - MS-2 Barcode Scanner - Communication error message improved when scanner is not found after initialization.
-
[GEX-2167] - GENESYS 30 USB comes unplugged - application will now keep report open.
-
[GEX-2182] - Calc Min Dose/Calc Max Dose no longer limited by Dose Range Max from the Calibration Configuration when using Correlation Ratios Dref:Dmax.
Bug fixes
-
[GEX-2164] - Saving Tote and/or Position in a report with no measurements now not allowed to avoid error.
-
[GEX-2169][GEX-2173][GEX-2178][GEX-2179]&[GEX-2180] - Dose calculations using higher order fits were giving errors for high values (affected FWT dosimeter users only). This was fixed. No calculations for 4.100 or higher absorbance values will be allowed. Also, extensive unit testing was added to verify the dose calculations every time a software build is compiled. Unit testing completed using all orders of fit that the software can handle over the full range of use of all compatable dosimeter types.
-
[GEX-2177] - CalibrationViewModel: ValidatableModel exceptions causing app to become unresponsive
-
[GEX-2186] - Calibration importing date formats allowable was expanded.
-
[GEX-2208] - The Edit ID features were available for use after a user processed a Report. When the user navigated away the issue was corrected. This fixes the issue completely.
Documentation
An updated User Guide - GEX Doc #100-266 rev H. is released.