Bug
(DCTRL-694) Fixed reread functionality for limited number of rereads. In versions 2.0.0, 2.0.1 and 2.0.2, if a user has configured ‘limit number of rereads’ for a Reread strategy, the software does not enforce a limit. Bug does not affect earlier versions 1.x.
(DCTRL-695) Fixed export to Excel error bug. Prior to version 2.0.3, if a dosimetry report contains ALL dosimeters with doses that are under or over the calibration dose range, there is an error message exporting the report to Excel.
Known Anomalies:
(DOSECTRL-239) Audit Trail displays incorrectly for statistical rereads. The individual measurement results should display with the statistical reread event. Currently, the Audit log displays each measurement as a unique event and incorrectly lists the statistical reread dose.
(DOSECTRL-80) If the user has ‘Edit Thickness’ permission enabled in the Dosimeter Type configuration, the user with permission is allowed to change the dosimeter thickness on the measurement screen.
(DOSECTRL-81) While the modification of the Calibration configuration End Date event is recorded accurately in the audit trail, the event is identified as GxP event and should have the user enter a comment to be recorded in the audit trail.
(DOSECTRL-82) Audit trail should record a change to the report's storage location.