DoseControl Software

Release Notes - version 2.0.3 (hotfix)

Bug

(DCTRL-694) Fixed reread functionality for limited number of rereads. In versions 2.0.0, 2.0.1 and 2.0.2, if a user has configured ‘limit number of rereads’ for a Reread strategy, the software does not enforce a limit. Bug does not affect earlier versions 1.x.

(DCTRL-695) Fixed export to Excel error bug. Prior to version 2.0.3, if a dosimetry report contains ALL dosimeters with doses that are under or over the calibration dose range, there is an error message exporting the report to Excel.

Known Limitations (non-critical):

Dev-035 If the user has ‘Edit Thickness’ permission enabled in the Dosimeter Type configuration, the user with permission is allowed to change the dosimeter thickness on the measurement screen.

Dev-036 While the modification of the Calibration configuration End Date event is recorded accurately in the audit trail, the event is identified as GxP event and should have the user enter a comment to be recorded in the audit trail.

Dev-037 Audit trail should record a change to the report's storage location.