Release Date: December 7, 2019
A summary of the highlights in this release.
New Features
[DC-2274] - Protect all PDF's with password to discourage editing with PDF editors
Improvements
This release contains the following improvements:
[DC-2285] - Add Reason to New Renamed Dosimeter Record in Readings table: reason was included with old name but not renamed dosimeter.
[DC-2286] - Audit Entries: Add flag to Readings table when barcode scan fails and user enters manually.
Bug fixes
[DC-2219] - Fix to Allow International Date Time Formats (database, code, and migration scripts)
[DC-2228] - MS-2 Barcode Scanner - Batch ID Verification Errors: 1 line of logic code added to prevent multiplicity of open scanners
[DC-2276] - Fix UI Display of International Date Time Formats
Known Anomalies:
(DOSECTRL-239) Audit Trail displays incorrectly for statistical rereads. The individual measurement results should display with the statistical reread event. Currently, the Audit log displays each measurement as a unique event and incorrectly lists the statistical reread dose.
(DOSECTRL-80) If the user has ‘Edit Thickness’ permission enabled in the Dosimeter Type configuration, the user with permission is allowed to change the dosimeter thickness on the measurement screen.