Bug

DCTRL-671 For systems that have the integrated barcode scanner, this bug fix creates a pop-up error message telling the user they have scanned a dosimeter ID that has a Batch ID that does not match the configured, active Batch in the system. In previous versions, the application would not take a measurement and crash when measuring a dosimeter using an integrated barcode scanner with an incorrect batch . Minor bug - no measurement is taken.

DCTRL-681 For systems that have a custom-PDF report that includes a re-read comment and re-read glyph, this bug fix places the reread glyph and comment on subsequent versions of the report. In previous versions, the PDF reports did not include reread glyphs or comments in later versions of a report. Minor bug affecting specific clients.

Known Limitations (non-critical):

Dev-035 If the user has ‘Edit Thickness’ permission enabled in the Dosimeter Type configuration, the user with permission is allowed to change the dosimeter thickness on the measurement screen.

Dev-036 While the modification of the Calibration configuration End Date event is recorded accurately in the audit trail, the event is identified as GxP event and should have the user enter a comment to be recorded in the audit trail.

Dev-037 Audit trail should record a change to the report's storage location.