Error Message |
Error Type |
Source |
|
Absorbance value: 0 out of reading range. You can take a new reading or skip. |
calculation integrity |
Calcuation integrity. Dosimeter measured absorbance is 0 or below 0 value |
|
Absorbance value: [abs value] out of reading range. You can take a new reading or skip. |
calculation integrity |
Dosimeter measured absorbance is not calculated because it is outside the dosimetry system calibration range |
|
Absorbance value: [below 0.000] out of reading range. You can take a new reading or skip. |
calculation integrity |
Dosimeter measured absorbance is 0 or below 0 value |
|
The identifier you provided is already in use. |
data integrity |
Data integrity. Duplicate item identifier - configuration name, dosimeter ID, report ID |
|
There is no active or valid calibration on this workstation for the Reader, Batch and Pathway specified. Unable to Proceed. |
data integrity |
Data integrity. Dosimetry system calibration is not configured and active in system for the associated spectrophotometer, pathway and dosimeter type. |
|
The reader used for this report does not match the reader currently attached to your workstation. Measurement functions are disabled. |
data integrity |
Data integrity. Spectrophotometer serial number does not match connected device. |
|
There is no active calibration on this workstation for the Reader, Batch and Pathway specified. Measurement functions have been disabled. |
data integrity |
Data integrity. Dosimetry system calibration is not active in system. |
|
There is no active or valid calibration on this workstation for the Reader, Batch and Pathway specified. In the Setup Calibrations section of DoseControl, please ensure that there is a calibration marked as active that matches the selected Reader, Batch and Pathway, and that calibration has correct Start and End Dates. Measurement functions have been disabled. |
data integrity |
Data integrity. Dosimetry system calibration is not active in system. |
|
The Dosimeter ID [scanned dosimeter ID] is used in another report (Report ID: [report ID]). You cannot use that ID again. Enter a different dosimeter ID. |
data integrity |
Data integrity. Dosimetry ID must be unique in system. |
|
The selected reader is overdue for calibration. Measurement functions have been disabled. |
data integrity |
Data integrity. Spectrophotometer is not valid for use; spectrophotometer calibration date expired. |
|
The selected reader is overdue for calibration. Unable to proceed. Measurement functions are disabled. |
data integrity |
Data integrity. Spectrophotometer is not valid for use; spectrophotometer calibration date expired. |
|
The selected reader is overdue for calibration. Unable to proceed. |
data integrity |
Data integrity. Spectrophotometer is not valid for use; spectrophotometer calibration date expired. |
|
All absorbances for this reading have been measured. |
data integrity |
Data integrity. Number of rereads allowed for a Dosimeter ID is exceeded. |
|
The measured dosimeter belongs to a different batch (EQ). The reading has been archived. |
data integrity |
Data integrity. The system's active Dosimeter batch does not match the scanned dosimeter batch/ID |
|
Command:GetWavelength N/A |
data integrity |
Data integrity. Software cannot verify the wavelength setting of the connected spectrophotometer. Problem with spectrophotometer connection and communication |
|
Command:SetWavelength N/A |
data integrity |
Data integrity. Software cannot verify the wavelength setting of the connected spectrophotometer. Problem with spectrophotometer connection and communication |
|
The reading's barcode (Zeroing can be performed inbetween each statistical Re-Read.) doesn't match the dosimeter id. |
data integrity |
Data integrity. The scanned Dosimeter ID does not match the expected Dosimeter ID during a reread session |
|
The reading's barcode (Test re-read 2) doesn't match the dosimeter id. |
data integrity |
Data integrity. The scanned Dosimeter ID does not match the expected Dosimeter ID during a reread session |
|
User tried to remove a field set that's in use. |
data integrity |
Data integrity. User cannot delete a configuration item used in a report or to make a measurement |
|
The measured dosimeter belongs to a different batch (CG). The reading has been archived. |
data integrity |
Data integrity. The system's active Dosimeter batch does not match the scanned dosimeter batch/ID |
|
There is no calibration configured for the selected reader, pathway and dosimeter type. Please select a different reader. |
data integrity |
Data integrity. Dosimetry system calibration is not configured and active in system for the associated spectrophotometer, pathway and dosimeter type. |
|
No Active or Valid Calibration on this workstation for the Reader, Batch and Pathway specified. Measurement functions disabled. |
data integrity |
Data integrity. The software cannot verify an active, valid spectrophotometer/dosimetry system combination. The spectrophotometer is not active, associated with the active dosimetry system Calibration. The spectrophotometer is not valid (reader's calibration end date expired) |
|
Printing failed. |
general error |
General error. Dosimeter report output (PDF or Excel) did not print |
|
Taking a reading failed. |
database connection error general error |
General measurement failure. Spectrophotometer communication error. Database connection error. |
|
Loading reader settings failed. |
database connection error general error |
General error. Software cannot load spectrophotometer configuration information due to database connection error |
|
Loading the measure screen data failed. |
database connection error general error |
General measurement failure. Spectrophotometer communication error. Database connection error. |
|
Processing failed. There was an error processing this report. Please try again. Technical information:"Sequence contains no elements" |
database connection error general error |
General report failure when user clicks Process report. Missing required report fields. Database connection error. |
|
Loading reports failed. |
database connection error general error |
General report screen loading failure. Database connection error. |
|
Saving fieldsets failed |
database connection error general error |
General software configuration failure due to issue with database connection. |
|
Loading audit entries failed. |
database connection error general error |
General audit trail screen laoding failure. Database connection error. |
|
Adjusting absorbances failed. |
database connection error general error |
General report screen loading failure. Database connection error. |
|
Adjusting absorbances failed. An error occurred while update the entries. See the inner exception for details. |
database connection error general error |
General measurement failure. Spectrophotometer communication error. Database connection error. |
|
Processing failed: Object reference not set to instance of an object. |
database connection error general error |
General report failure when user clicks Process report. Report screen has not fully loaded due to database connection speed. Database connection error. Missing required report fields or data type mismatch in report fields. |
|
Loading barcode scanner failed. |
integrated barcode scanner comm error |
Internal barcode scanner, part of Reader configuration, communication error. Software not able to communication with the scanner |
|
Matrix 120 not found on any active COM ports. Ensure the barcode scanner is connected to this PC. |
integrated barcode scanner comm error |
Internal barcode scanner, part of Reader configuration, communication error. Software not able to communication with the scanner |
|
You don’t have permission to save in this location |
PC error |
PC error. Windows permission issue for the logged in user is preventing file saving to the report storage location, as configured in the software |
|
Printing failed. The system cannot find the file specified. |
PC error |
PC error. Windows permission issue for the logged in user is preventing file saving to the report storage location, as configured in the software |
|
We found a problem with some of the content in ‘your file name.xlsx’. Do you want us to try and recover as much as we can? |
PC error |
PC error. Excel template corruption. |
|
System.Exception: Missing file exporter modules: [Module name] |
PC error |
Missing a report module in GEX application installation files on PC |
|
Port driver error in Read() Source: System Message: The I/O operation has been aborted because of either a thread exit or an application request. |
Reader communication error |
Spectrophotometer communication error. Software not able to communicate with the spectrophotometer |
|
Reader timed out |
Reader communication error |
Spectrophotometer communication error. Software not able to communicate with the spectrophotometer |
|
Command:GetReaderStatus st |
Reader communication error |
Spectrophotometer communication error. Software not able to communicate with the spectrophotometer |
|
Port driver error in OpenConnection() Source: System Message: Access to the port 'COM5' is denied. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. |
|
Communicating with the reader failed. Ensure that the reader is connected and the correct COM port is configured. |
Reader communication error |
Spectrophotometer communication error. Software not able to communication with the spectrophotometer |
|
Reader is not initialized |
Reader communication error |
Spectrophotometer communication error. Software not able to communication with the spectrophotometer |
|
Port driver error in OpenConnection() Source: System Message: Access to the port '[COM NUMBER]' is denied. |
Reader communication error |
Spectrophotometer communication error. Software not able to communicate with the spectrophotometer |
|
Port driver error in OpenConnection() Source: System Message: The port '[COM number]' does not exist. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. Problem with PC COM port functionality. |
|
Port driver error in Write() Source: System Message: A device attached to the system is not functioning. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. Problem with PC COM port functionality. |
|
Port driver error in Read() Source: System Message: The BaseStream is only available when the port is open. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. Problem with PC COM port functionality. |
|
Port driver error in Read() Source: GEX.Common Message: Port driver error in Flush() Source: System Message: The BaseStream is only available when the port is open. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. Problem with PC COM port functionality. |
|
Taking a reading failed. Ensure that the reader is connected and the correct COM port is configured. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. Problem with PC COM port functionality. |
|
Port driver error in OpenConnection() Source: System Message: Access to the port 'COM3' is denied. |
Reader communication error |
Spectrophotometer disconnected from system or cannot be found. Problem with PC COM port functionality. |
|
Connection failed to initialize. |
Reader communication error |
Spectrophotometer communication error. Software not able to communicate with the spectrophotometer |
|
You need to have the Reread role. |
user access and permissions |
User access and permissions |
|
Login failed. |
user access and permissions |
user login |
|
LDAP login failed. |
user access and permissions |
user login |
|
Loading [software screen] failed. The EXECUTE permission was denied on the object [database object], database [connected database name], schema 'dbo.' |
database connection error general error |
Database connection error. Database connection string is incorrect, as configured in software. Database/server parameters have changed. SQL username/password in connection string is incorrect or disabled. |